Ensure the CPAP Machine is set up properly according to its manual.
Connect the male end of the Oxygen Enrichment Adapter to the open female end of the CPAP tube.
Connect the male end of a Tee Connector to the female end of the Oxygen Enrichment Adapter.
Connect the female end of a Viral Filter to the male top of the Tee Connector. Moreover, connect a PEEP Valve on the female end of the Tee Connector.
Finally, connect a Non-Vented Mask to the Viral Filter.
Figure 1: CPAP circuit with one viral filter and one PEEP valve
Video 1: CPAP Version 1 Assembly
CPAP Vesion 1 Components
Specification
Figure
CPAP Machine
-
Non-Vented Mask
-
Tee Connector
22 mm ID x 15 mm ID x 22 mm OD
AMBU Peep Valve
22 mm OD
Oxygen-Enrichment Adapter
-
Viral Filter
22mm I.D. x 15mm ID / 22mm OD
CPAP Version 2
Ensure the CPAP Machine is set up properly according to its manual.
Connect the male end of the Oxygen Enrichment Adapter to the open female end of the CPAP Tube.
Connect two Tee Connectors (TC1 and TC2) in a row (female end of TC1 connected to the male end of TC2). Next, attach the male end of TC1 to the female end of the Oxygen Enrichment Adapter.
On the female end of TC2 attach the male end of an Elbow Connector.
Connect the female end of a Viral Filter on the male top of TC2.
Connect a Female Coupler on the male top of TC1.
Now connect one PEEP Valve to the Female Coupler on TC1 and another PEEP Valve to the Elbow Connector.
Finally, attach the Non-Vented mask to the Viral Filter.
Figure 2: CPAP circuit with one viral filter and two PEEP valves
Video 2: CPAP Version 2 Assembly
CPAP Version 2 Components
Specification
Figure
CPAP Machine
-
Non-Vented Mask
-
Tee Connector 1 (TC1)
22 mm ID x 15 mm ID x 22 mm OD
Tee Connector 2 (TC2)
22 mm ID x 15 mm ID x 22 mm OD
Female Coupler
22mm ID x 22mm ID
Elbow Connector
22 mm OD / 15 mm ID x 15 mm OD
AMBU Peep Valve 1
22 mm OD
AMBU Peep Valve 2
22 mm OD
Oxygen-Enrichment Adapter
-
Viral Filter
22mm I.D. x 15mm ID / 22mm OD
CPAP Version 3
Ensure the CPAP Machine is set up properly according to its manual.
Connect the male end of the Oxygen Enrichment Adapter to the open female end of the CPAP Tube.
Connect the male end of a Tee Connector (TC1) to the female end of the Oxygen Enrichment Adapter.
On the female end of TC1 connect the male end of a Viral Filter (VF1).
Connect the male end of a Tee Connector (TC2) to the female end of VF1. Then, on the female end of TC2, connect the male end of another Viral Filter (VF2).
Connect the male end of an Elbow connector to the female end of VF2.
Attach a Female Coupler to TC1.
Attach one PEEP Valve to the Female Coupler on TC1 and another PEEP Valve to the Elbow Connector.
Finally attach the NON-Vented mask to the top of TC2.
Figure 3: CPAP circuit with two viral filters
Video 3: CPAP Version 3 Assembly
CPAP Version 3 Components
Specification
Figure
CPAP Machine
-
Non-Vented Mask
-
Tee Connector 1 (TC1)
22 mm ID x 15 mm ID x 22 mm OD
Tee Connector 2 (TC2)
22 mm ID x 15 mm ID x 22 mm OD
Female Coupler
22mm ID x 22mm ID
Elbow Connector
22 mm OD / 15 mm ID x 15 mm OD
AMBU Peep Valve 1
22 mm OD
AMBU Peep Valve 2
22 mm OD
Oxygen-Enrichment Adapter
-
Viral Filter (VF1)
22mm I.D. x 15mm ID / 22mm OD
Viral Filter (VF2)
22mm I.D. x 15mm ID / 22mm OD
BiPAP
Ensure the BiPAP Machine is set up properly according to its manual.
Connect the male end of the Oxygen Enrichment Adapter to the open female end of the BiPAP tube.
Connect the male end of a Tee Connector to the female end of the Oxygen Enrichment Adapter.
Connect the female end of a Viral Filter to the male top of the Tee Connector. Moreover, connect a PEEP Valve on the female end of the Tee Connector.
Finally, connect a Non-Vented Mask to the Viral Filter.